OxyTrans 4 – 20 mA Loop Powered two wire
Oxygen Transmitter
Two Wire Process Oxygenanalyzer
The Model OxyTrans is simple, accurate andeconomic measurement of Oxygen at traceand % levels.
A simple Display and one point calibration allow easy and fast start-up of the instrument.
optional 1/2 G Ex ia IIC T4 (ATEX)
One User specific measuring range
The unit will offer with ranges between
0-10 ppm and 0 – 100% of oxygen.
Sensor
The OxyTrans uses a special fuel cell to measure the oxygen concentration. The sensor meets the industrial requirements for accuracy, sensitivity, easy to use and operating life.
Calibration
The calibration of the instrument for trace oxygen measurements in gas, should be done with a calibration gas. The concentration can be chosn freely within the measuring range. In the percent range the unit can calibrated with air.
Features:
compact
inexpensive
Sensor with long operating life
IP65 enclosure
reverse voltage protection and temperature compensation
Flow-through measuring cell inStainless Steel
• The measuring cell is modular and is made of stainless steel
• A defective measuring cell can be repaired by replacing the defective part only, rather than the complete unit